User's Manual

Table Of Contents
Monitoring SpO2 (Optional) Understanding the SpO2 Display
TMS60 Operator’s Manual 8 - 11
8.7.2 About the SpO
2
Digital Area
The SpO
2
digital area displays in units of % with a resolution of 1%.
The SpO
2
PR value displays in units of bpm with a resolution of 1 bpm.
Displays Masimo PI resolution as 0.01 when the PI value is smaller than 10%.
Displays Masimo PI resolution as 0.1 when the PI value is greater than or equal to
10%.
If the SpO
2
measurement or PR is invalid, “- - -” displays in place of digits.
8.7.3 SpO
2
Waveform Area
The SpO
2
waveform area displays:
1. Signal Indicator Quality (SIQ)
2. Area name
3. Pleth waveform
Masimo SpO
2
waveform area (SIQ enabled)
Nonin SpO
2
waveform area
8.7.4 About the SpO
2
Waveform Area
Displays in the configured SpO
2
color.
Provides scrolling, real-time waveform data.
Scrolls the waveform in the configured sweep speed.
Automatically scales the SpO
2
waveform data area to maximize the vertical height
of the Pleth waveform for the data range.
If using Masimo SpO
2
, the Signal Quality Index (SIQ) will display below the
waveform if enabled.
3
2
1