Monitoring SpO2 (Optional) SpO2 Measurement
TMS60 Operator’s Manual 8 - 9
WARNING
When equipped with Masimo SpO
2
module, use only Masimo SpO
2
sensors specified in this manual. Use of other SpO
2
sensors may cause
improper oximeter performance.
When equipped with Nonin SpO
2
module, use only Nonin SpO
2
sensors
specified in this manual. Use of other SpO
2
sensors may cause improper
oximeter performance.
Do not disconnect the Nonin Spo
2
sensor connector from the Nonin
SpO
2
module during defibrillation.
If the sensor is too tight because the application site is too large or
becomes too large due to edema, excessive pressure for prolonged
periods may result in venous congestion distal from the application site,
leading to interstitial edema and tissue ischemia.
CAUTION
Many patients suffer from poor peripheral perfusion due to
hypothermia, hypovolemia, severe vasoconstriction, reduced cardiac
output, etc. These symptoms may cause a loss in vital sign readings.