TMS60 Operator’s Manual 7 - 11
7.4 Changing the ECG Settings
You can change the ECG settings from the ECG menu.
7.4.1 Configuring the ECG Setup
Enter the ECG menu in either of the following ways:
On the main screen, tap the HR digital area or ECG waveform area to enter the
[ECG] menu.
In the main menu, tap [Parameter Setup] [ECG] to enter the [ECG] menu.
1. In the [Setup] section of the [ECG] menu, select the options described in the fol-
lowing table.
NOTE
When [Paced] is set to [Yes], the system does not detect PVC-related
arrhythmia (including PVCs) resulting from pacemaker but still analyzes
the normal QRS complex.
Options Description Settings*
Lead Placement Enters the [Lead Placement] window. Refer to "Checking the Lead
Placement" on page 7 - 9 for
details.
Cable Type Selects the current ECG leadwire type. Auto, 3 Lead, 5 Lead
Refer to "ECG Leadwire Types" on
page 7 - 12 for details.
Smart Lead (Monitored
Lead)
When [Cable Type] is set to [Auto], the option displays [Smart Lead].
Drag the swtich to right or left to enable or disable the Smart Lead function.
When [Cable Type] is set to [3 Lead], the option displays [Monitored Lead].
Refer to "ECG Leadwire Types" on page 7 - 12 for details.
* The factory default settings are in bold.