TMS60 Operator’s Manual 7 - 7
7.3.4.2 Standard 5-Leadwire Electrode Placement
A 5-wire lead set can monitor seven ECG vectors (I, II, III, aVR, aVL, aVF, and V) simultane-
ously. The recommended 5-wire lead placement is as follows:
5-wire lead placement (AHA)
5-wire lead placement (IEC)
Place the RA (white) electrode under the
patient’s right clavicle, at the mid-
clavicular line within the rib cage frame.
Place the LA (black) electrode under the
patient’s left clavicle, at the mid-clavicular
line within the rib cage frame.
Place the LL (red) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Place the RL (green) electrode on the
patient’s lower right abdomen within the
rib cage frame.
Place the V (brown) electrode in the V-
lead position as shown in the figure or
decided by the clinician.
Place the R (red) electrode under the
patient’s right clavicle, at the mid-
clavicular line within the rib cage frame.
Place the L (yellow) electrode under the
patients left clavicle, at the mid-clavicular
line within the rib cage frame.
Place the F (green) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Place the N (black) electrode on the
patient’s lower right abdomen within the
rib cage frame.
Place the C (white) electrode in the C-lead
position as shown in the figure or
decided by the clinician.