User's Manual

Table Of Contents
Alarm Indicators Alarms
6 - 4 TMS60 Operator’s Manual
6.4.3 Alarm Messages
When a technical alarm occurs on the TD60 screen, the alarm message appears in the
message area. The background color of the alarm message and the asterisk symbols (*)
before the alarm message are designed to indicate the alarm level.
6.4.4 Alarm Status Symbols
The TD60 still uses the following symbols indicating the alarm status:
: indicates the technical alarm audio is turned off.
: indicates the alarm system is reset.
High level alarms:
Medium level alarms:
Low level alarms:
high-pitched triple beep
double beep
low-pitched single beep
NOTE
When multiple technical alarms of different levels occur simultaneously,
the TD60 selects the alarm of the highest level to light the alarm light
and sound alarms accordingly, while all the alarm messages scroll in the
message area on the top of the screen.
Alarms Background color Asterisk symbols (*)
High level alarms red ***
Medium level alarms yellow **
Low level alarms cyan *