User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Measurement Specifications Product Specifications
A - 10 TMS60 Operator’s Manual
A.8 Measurement Specifications
A.8.1 ECG
ECG
Standard
Meet standards of IEC60601-2-27 and IEC60601-2-25
Leadset
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
Automatic 3/5 lead recognition
ECG standard
AHA, IEC
Sweep speed
6.25 mm/s, 12.5 mm/s, 25 mm/s
Accuracy: ± 10%
Display Sensitivity
1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/
mV (×1), 20 mm/mV (×2), 40 mm/mV (×4)
Accuracy: ± 5%
Input signal range
±8 mV (peak-to-peak value)
Calibration signal
1 mV (peak-to-peak value)
Accuracy: ± 5%
Differential input impedance
≥ 5 M
Bandwidth (-3dB)
Monitor mode: 0.5 Hz to 40 Hz
ST mode: 0.05 Hz to 40 Hz
Common mode rejection ratio
(with Notch off)
Monitor mode: > 105 dB
ST mode: > 105 dB
Notch
50/60 Hz, rejection capacity ≥ 20 dB
Polarizing voltage scope
± 500 mV
Noise
≤ 30 µV (p-v RTI)
Baseline recovery time
< 5 s (after defibrillation)
Direct current leakage
Input electrode: < 0. 1µA
Drive electrode: <1µA
Electrode polarization recovery time
< 10 s