Maintenance and Testing Schedule Maintenance
14 - 4 TMS60 Operator’s Manual
14.4 Maintenance and Testing Schedule
The following maintenance and tests, except for visual inspection, power on test, and
battery check, shall be carried out by the service personnel only. Contact your service
personnel if any maintenance is required. Make sure to clean and disinfect the equip-
ment before any test and maintenance.
14.5 Checking the System Information
To view the information about the device, radio frequency (RF), battery, MPAN, and sys-
tem statistics, you can go to the main menu [System Info].
14.6 Verifying the ECG at the TD60
The ECG signal may be inaccurate due to hardware or software problems. As a result, the
ECG wave amplitude becomes greater or smaller. To verify the ECG waveform amplitude:
1. In the main menu, tap [Maintenance].
2. Input the maintenance passcode.
3. Tap [Accept].
4. In the [Maintenance] menu, tap [Others].
5. Enable [Calibrate ECG].
A square wave appears on the screen and the message [ECG Calibrating] is dis-
played.
Check/Maintenance Item Recommended Frequency
Visual inspection When first installed or reinstalled.
ECG test and verification Performance test 1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of rele-
vant module.
3. Once every two years.
Verification
SpO
2
test
Power on test 1. When first installed or reinstalled.
2. Following any maintenance or the replace-
ment of any main unit parts.
Battery check Functionality test 1. When first installed.
2. Whenever the battery is replaced.
Performance test When the battery run time reduced significantly.