User's Manual

Table Of Contents
Cleaning and Disinfecting Cleaning the Battery and Battery Compartment
TMS60 Operator’s Manual 13 - 5
13.5 Cleaning the Battery and Battery Compartment
Clean the exterior surface of the lithium-ion battery pack and AA battery tray
before and after each new patient.
Wipe the lithium-ion battery pack, AA battery tray and the battery compartment
using a soft cloth with mild soap and water solution. Use caution to avoid the
battery connector.
Allow the lithium-ion battery pack, AA battery tray and the battery compartment
completely dry before using.
Refer to the cleaning procedure described in "Cleaning of the TD60" on page 13 - 3.
13.6 Disinfection
Disinfect the TD60 and accessories as required in your hospital’s servicing schedule.
Cleaning before disinfecting is recommended.
The recommended disinfectants are:
70% isopropyl alcohol
10% sodium hypocholride (bleach) solution
3% hydrogen peroxide
Virkon
Super Sani-cloth (0.5% Quaternary ammonium chloride and 55% Isopropyl
alcohol)
50% propyl alcohol (1-propyl alcohol)
70% ethanol
13.7 Sterilization
Sterilization is not recommended for this equipment, related products, accessories or
supplies unless otherwise indicated in the Operating Instructions that accompany the
accessories or supplies.
CAUTION
Never immerse the lithium-ion battery pack, AA battery tray and AA
batteries. Do not clean them with harsh chemicals such as acetone or
nondiluted bleach.