User's Manual

Table Of Contents
Technical Alarm Messages at the TD60 Troubleshooting
12 - 6 TMS60 Operator’s Manual
ECG ECG
Noise
Low A The ECG signal is noisy. Check for any
possible sources of
signal noise around
the cable and
electrode, and check
the patient for great
motion.
ECG
Cable
Type
Error
Low B Connect wrong ECG
leadwire.
Reconnect the 3 or 5
lead ECG leadwire.
HR
Overrang
e
Low A HR exceeds the
measurement limit.
Contact your
medical pesonnel.
SpO
2
SpO
2
Sensor
Off
Low* B The SpO
2
sensor has
become detached from
the patient or the
module.
There is a fault with the
SpO
2
sensor.
An unspecified SpO
2
sensor has been used.
Check the sensor
application site and
the sensor type, and
make sure the
sensor is not
damaged.
Reconnect the
sensor or use a new
sensor.
SpO
2
Sensor
Fault
Low C
SpO
2
No
Sensor
Low B
SpO
2
Module
Error
High A
SpO
2
Too
Much
Light
Low A There is too much light
on the SpO
2
sensor.
Move the sensor to a
place with lower
level of ambient
light or cover the
sensor to minimize
the ambient light.
SpO
2
No
Pulse
Low A SpO
2
sensor failed to
obtain pulse signal.
Move the sensor to a
site with better
perfusion.
SpO
2
Unplugg
ed
Low C SpO
2
module connector
is disconnected from the
TD60.
Reconnect the SpO
2
module to the TD60.
PR
Overrang
e
Low A The measured PR value
exceeds the
measurement range.
Contact Mindray or
your service
personnel.
Measurement
Alarm
message
Alarm
level
Alarm
Indication
Possible cause Solution
* The asterisk (*) means the alarm level is configurable.
** XX is the ECG lead name.