User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Configuring the TD60 Maintenance Menu
TMS60 Operator’s Manual 10 - 7
3. Tap to exit the [Defaults] menu.
10.2.6 Transferring a Configuration
The TD60 is capable of configuring multiple devices through one export operation via a
wireless interface.
To transfer the configuration from the TD60 unit to an external device:
1. In the [Maintenance] menu of an external device, tap [Defaults] → [Import
Device Settings] to enter the settings import mode.
2. In the [Maintenance] menu of the TD60 unit, tap [Defaults] → [Export Device
Settings] to enter the [Export Device Settings] menu.
The discovered external devices are listed in the [Export Device Settings] menu.
3. Select the desired external devices by tapping the check box.
4. Tap [Export] to start exporting the TD60 unit configuration.
The selected external devices screen will shortly display the “Downloading device
settings” message.
■ On the external devices if an import is successful, the external device will display
the “Download complete.” message, and after 10 seconds return to the [Defaults]
menu. On the TD60 unit, the status for the external device will display the
“Complete” message.
■ On the external device if the import fails, the external device will display the
“Import attempt failed.” message. On the TD60 unit, the status for the failed
external device will display the “Failure” message and remain in the list.
The user has two options when a failure occurs:
◆ To stop the import from the external device, tap the icon to exit the
[Import Device Settings].
Import Device Settings Copies the settings from the external device to theTD60 unit. Refer to "Trans-
ferring a Configuration" on page 10 - 7 for details.
Restore Factory Defaults Allows an authorized user to reestablish the original database power up
settings to factory default values.
Options Description
WARNING
• Do not power off the devices during the download process.