User's Manual

Table Of Contents
Arrhythmia Monitoring Monitoring with the TD60 at the CS
9 - 32 TMS60 Operator’s Manual
Minimum QRS Threshold menu
The current waveform displays the data of previous eight seconds. Use the buttons
described in the following table as desired.
2. Select the [OK] or [Cancel] button.
Once the threshold is in the desired position, select the [OK] button to save
the settings and close the menu.
To quit adjusting the threshold, select the [Cancel] button to close the menu
without saving the settings.
Buttons Description Settings*
Gain Select the desired ECG waveform size. X1, X2, X4
Refresh Displays the real-time waveform.
Default Automatically sets the default threshold: 0.16 mV.
or
Manually adjust the minimum QRS detection threshold.
Select the button to move the threshold line above the P wave, or select the
button to move the threshold line down closer to the P wave.
* The factory default settings are in bold.