User's Manual

Table Of Contents
Arrhythmia Monitoring Monitoring with the TD60 at the CS
9 - 30 TMS60 Operator’s Manual
9.6.3 Arrhythmia Settings
The following table lists the arrhythmia settings in the [Arrhythmia Analysis] section of
the [ECG] tab.
9.6.4 Relearning
A relearn can be done for arrhythmia, ST analysis, or for both simultaneously.
The TD60 initiates the learning process for ST or Arrhythmia analysis after any of the fol-
lowing:
TD60 power-up
Return to normal monitoring from the Standby mode
Enabling ST or Arrhythmia analysis
The lead has been changed in ECG 1 waveform (3 lead only)
Patient size changes
Selecting the Relearn button in the [Arrhythmia Analysis] section
Selecting the Relearn button in the [Arrhythmia Analysis] section is recommended
after one or more of the following:
ECG electrodes have been repositioned
Eight hours have passed since the last relearn
Significant changes occurred to the patient QRS complex
Options or Buttons Description Settings*
Arrhythmia Alarms Configures the arrhythmia alarm settings. Refer to "Factory Default Arrhyth-
mia Alarm Settings" on page 9 - 6
for details.
Relearn
Enables an
arrhythmia relearning.
Refer to "Relearning" on page 9 -
30 for details.
QRS Threshold Settings Configures the QRS threshold. Refer to "Configuring the QRS
Threshold" on page 9 - 31 for
details.
NOTE
ST Analysis must be turned on before it will relearn.