9 - 24 TMS60 Operator’s Manual
9.5.3 Adjusting ST Measurement Points
The ISO and ST points need to be adjusted when you start monitoring and if the
patient’s heart rate or ECG morphology changes significantly. Exceptional QRS com-
plexes are not considered for ST-segment analysis.
1. In the [ST Analysis] section of the [ECG] tab, click [Define ST Point].
The [Define ST Point] menu displays, as shown in the following figure.
Define ST point menu
2. Adjust the parameter using the buttons described in the following table.
WARNING
Always make sure that the ST measurement points are appropriate for
your patient.
Buttons Description Settings*
Lead Select the desired ECG lead. I, II, III, aVR, aVL, aVF, V
ST (ms) Based on the ms setting selected, moves
the ST point further or closer to the J point
in the ST template.
J+40, J+60, J+80, J+60/80
* The factory default settings are in bold.