9.4.3 QT View Menu (Only for Mindray ECG Algorithm)
In the [QT View] menu, you can view a snapshot of the real-time wave and to verify that
the QT algorithm detects correct Q and T points.
The [QT View] menu displays, as shown in the following figure:
The current waveform and parameter values display in green.
The template waveform and parameter values display in yellow.
The Q and T points are marked with a vertical line.
The QTc value is equal to the current QTc value minus the template QTc
value.
QT View menu
Using the buttons described in the following table as desired.
Buttons Description Settings*
Lead Select the desired lead to display on the
[QT View] menu screen.
II, I, III, aVR, aVL, aVF, V
Set as Ref. Replaces the template waveform and QT/
QTc values with the current waveform and
QT/QTc values.
None
The QT template updated time displays at the bottom of the screen.
Print Prints the template and current wave-
forms, and QT/QTc values for all leads.
None
Exit Closes the [QT View] Menu. None
* The factory default settings are in bold.