User's Manual

Table Of Contents
Monitoring with the TD60 at the CS QT Monitoring
TMS60 Operator’s Manual 9 - 19
QT Computational
Formula
Configures the QTc formula used.
Bazett:
Fridericia:
Framingham:
Hodges:
QT Alarm Setup Configures the QT alarms.
For details about the QT alarms limits, refer
to "Factory Default Parameter Alarm Limits"
on page 9 - 3 for details.
For details about the QT alarms responses,
refer to "Factory Default Parameter Alarm
Settings" on page 9 - 5 for details.
For Mindray algorithm, you can configure the QTc and QTc alarm.
For Mortara algorithm, you can configure the QTc alarms.
QT View Select to display the [QT View]
menu.
Refer to "QT View Menu (Only for Mindray ECG
Algorithm)" on page 9 - 20 for details.
This button is only available for the Mindray ECG algorithm.
Options Description Settings*
* The factory default settings are in bold.
QTc QT
QTHR
60
-------------


12/
×=
QTc QT
QTHR
60
-------------


13/
×=
QTc QT
154 1
60
QTHR
-------------


×+=
QTc QT
1.75
QTHR
60()×+=