User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Physiological Alarms Monitoring with the TD60 at the CS
9 - 10 TMS60 Operator’s Manual
9.2.5 Arrhythmia Threshold Settings
When an arrhythmia violates its threshold, an alarm is triggered. For the Mortara algo-
rithm, the setting of asystole delay is related to arrhythmia relearn. When HR is less than
30 bpm, it is recommended to set asystole delay to 10 s.
9.2.5.1 Mindray ECG algorithm
Arrh. event Range or Option Default Step
Unit of
measure
PVCs High 1 to 100 10 1 minute
Pauses/min 1 to 15 8 1 N/A
Asys. Delay 3 to 10 4 1 second
Tachy High 60 to 300 Adult: 120
Pediatric: 160
5bpm
Brady Low 15 to 120 Adult: 50
Pediatric: 75
5bpm
Extreme Tachy 60 to 300 Adult: 160
Pediatric: 180
5bpm
Extreme Brady 15 to 120 Adult: 35
Pediatric: 50
5bpm
Multif. PVC’s Window 3 to 31 15 1 /min
Vtac Rate 100 to 200 130 5 bpm
Vtac PVC 3 to 99 6 1 beat
Pause Time 1.5, 2.0, and 2.5 2.0 N/A second
Vbrd Rate 15 to 60 40 5 bpm
Vbrd PVCs 3 to 99 5 1 beat