Monitoring with the TD60 at the CS Physiological Alarms
TMS60 Operator’s Manual 9 - 9
In addition, the activation state of some arrhythmias can be set as a whole with the fol-
lowing buttons that are at the bottom of the Arrhythmia Alarms tab.
Irr.Rhythm X
X
Off Off Off Off
Pauses/min X
X
X X On Off Off Off
Parameters
Alarm levels*
Activation
State
Factory default alarm
responses
High Medium Low Message
Record
on
Alarm
Print
on
Alarm
Paging
Switch
* X indicates available alarm level, — indicates alarm level not available
** The factory default settings are in bold.
NOTE
The priority of lethal arrhythmia alarms is always high. It is unchange-
able.
Button Description
Lethals Only Sets the lethal arrhythmia alarms to on and all non-lethal arrhythmia alarms to off.
All Alarm On Sets all arrhythmia alarms to on.
All Alarm Off Sets all arrhythmia alarms to off.
This button is enabled when the Lethal Arrh. OFF option in the Te le me tr y tab from
the Admin Setup menu is set to Enable.