Table Of Contents
Monitoring ECG Changing the ECG Settings
TMS60 Operator’s Manual 7 - 13
2. Tap to exit the ECG menu.
7.4.4 Configuring the Pacer
1. In the Pacer section of the ECG menu, tap the options described in the following
table.
2. Tap to exit the ECG menu.
Speed Selects the waveform sweep speed. 6.25 mm/s, 12.5 mm/s, 25 mm/s
Options Description Settings*
* The factory default settings are in bold.
Options Description Settings*
Paced Selects the paced status. Unspecified, No, Yes
Unspecified is only available for
the first time you set the paced sta-
tus.
Refer to "Checking the Paced Sta-
tus" on page 7 - 10 for details.
Markers Selects the pacer indicator.
Line
A 1 cm line shows above each ECG wave-
form each time the pace pulse is detected.
Dot
A 2 mm dot shows above each ECG wave-
form each time the pace pulse is detected.
Line, Dot, Off
Pacer Reject Selects whether or not to reject the pace
pulses.
On, Off
* The factory default settings are in bold.
NOTE
When Paced is set to Yes, the Markers and Pacer Rejection options can
be available.