User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- C FCC Compliance
- D Symbols and Abbreviations
- E Anomaly
Monitoring ECG Changing the ECG Settings
TMS60 Operator’s Manual 7 - 11
7.4 Changing the ECG Settings
You can change the ECG settings from the ECG menu.
7.4.1 Configuring the ECG Setup
Enter the ECG menu in either of the following ways:
■ On the main screen, tap the HR digital area or ECG waveform area to enter the ECG
menu.
■ In the main menu, tap Parameter Setup → ECG to enter the ECG menu.
1. In the Setup section of the ECG menu, select the options described in the follow-
ing table.
NOTE
• When Paced is set to Yes, the system does not detect PVC-related
arrhythmia (including PVCs) resulting from pacemaker but still analyzes
the normal QRS complex.
Options Description Settings*
Lead Placement Enters the Lead Placement window. Refer to "Checking the Lead
Placement" on page 7 - 9 for
details.
Cable Type Selects the current ECG leadwire type. Auto, 3 Lead, 5 Lead
Refer to "ECG Leadwire Types" on
page 7 - 12 for details.
Smart Lead (Monitored
Lead)
When Cable Type is set to Auto, the option displays Smart Lead.
Drag the swtich to right or left to enable or disable the Smart Lead function.
When Cable Type is set to 3 Lead, the option displays Monitored Lead.
Refer to "ECG Leadwire Types" on page 7 - 12 for details.