TMS60 Operator’s Manual 6 - 5
6.5 Configuring the Alarms
For details on configuring the technical alarm volume, refer to "Configuring the
Audio Volume" on page 4 - 5.
For details on configuring the TD60 technical alarm settings, refer to "Configuring
the Alarms Menu" on page 10 - 3.
For the CS alarm configurations, refer to the BeneVision Central Station Opera-
tor’s Manual (P/N 046-007960-00).
6.6 Resetting the Alarms
You can acknowledge the on-going alarms by resetting the alarms. After being reset the
alarm system can respond to a subsequent alarm condition.
When a technical alarm occurs, follow this procedure to reset the TD60 alarm system.
Press the key to enter the main menu, and then tap Alarm Reset from the
Commands section.
OR
1. Swipe your finger up at the bottom of the main screen to display the quick keys
area.
2. Tap the Alarm Reset quick key to reset the alarm system.
When the alarm system is reset, depending upon the technical alarm there are several
ways the alarm system may respond as follows:
The alarm sound will be silenced, the alarm light will continue to indicate the
alarm, a √ will appear before the alarm message. The symbol appears on the
top of the main screen.
The technical alarm will be changed to the prompt message, it will not longer
make sound or be indicated by the alarm light.
The technical alarms are cleared, there will be no alarm indications.
For details about the indications of technical alarms when the alarm system is reset, refer
to "Technical Alarm Messages at the TD60" on page 12 - 5.