User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- C FCC Compliance
- D Symbols and Abbreviations
- E Anomaly
General Product Description Antenna Array
TMS60 Operator’s Manual 2 - 5
2.6 Antenna Array
The antenna array must be installed and configured by Mindray authorized personnel.
For more details about the antenna array installation, calibration, and validation, refer to
the Telemetry Monitoring System Installation Guide (P/N 046-007624-00).
2.7 Telemetry Receiver (RC60)
The RC60 receives the data from the TD60 via the antenna array, and sends the data to
the CS for analysis and display.
For details about the general wireless communication problems, refer to "General Prob-
lems" on page 12 - 2.
For details about the frequency allocation and receiver connection, refer to the Te le me -
try Monitoring System Installation Guide (P/N 046-007624-00).
For details about the RC60, refer to the TMS60 Service Manual (P/N 046-007057-00).
2.8 Touch Screen Display
Move your finger on the touch screen display to operate the TD60. For details about the
supported touch gestures, refer to "Understanding the Screen Display Orientation" on
page 3 - 10.
WARNING
• Authorized Mindray personnel are required to confirm the coverage
area of the antenna array in the following situations:
◆ When the antenna array is initially installed.
◆ When the antenna array is modified.
◆ When the building construction is modified.
WARNING
• Do not operate the touch screen with water on the hand.