A - 16 TMS60 Operator’s Manual
Bland-Altman Plots of the Bias versus the Mean
A.8.2.3 Masimo SpO
2
Module
SpO
2
Measurement range
1% to 100%
Resolution
1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured with motion)
1% to 69%: Not specified.
Low Perfusion Conditions
Pulse amplitude: > 0.02%
Light penetration: > 5%
Low Perfusion Accuracy
2%
*SpO
2
accuracy was determined by testing on healthy adult volunteers in the range of 60% to 100% SpO
2
against a laboratory CO-Oximeter. SpO
2
accuracy was determined on 16 neonatal NICU patients ranging in age
from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over
range of 70-100% SaO
2
with a resultant accuracy of 2.9% SpO
2
.
The following table shows the accuracy (A
rms
) for the SpO
2
sensors in four kinds of SaO
2
ranges. For the Fitting
Curve of the SpO
2
sensors, refer to the "Fitting Curve for Masimo SpO2 Sensors" on page A - 18.