Product Specifications Measurement Specifications
TMS60 Operator’s Manual A - 13
Resolution
0.01 mV
Accuracy
-0.8 mV to +0.8 mV: ± 0.02 mV or ± 10%, whichever is greater
Beyond this range: Not specified
Arrhythmia Analysis
Arrhythmia analysis classifications
Mindray ECG algorithm Mortara ECG algorithm
Asystole, VFib/VTac, Vtac, Vent.
Brady, Extreme Tachy, Extreme
Brady, PVCs/min, Pauses/min, R
on T, Run PVCs, Couplet, Multif.
PVC, PVC, Bigeminy, Trigeminy,
Tachy, Brady, Pacer Not Pacing,
Pacer Not Capture, Missed Beat,
Nonsus. Vtac, Vent. Rhythm,
Pause, Irr.Rhythm, Afib
Asystole, VFib, Vtac, Extreme
Tachy, Extreme Brady, PVCs/
min, Pauses/min, R on T, Run
PVCs, Couplet, Multif. PVC,
Bigeminy, Trigeminy, Tachy,
Brady, Pacer Not Pacing, Pacer
Not Capture, Vent. Rhythm,
Pause, Irr.Rhythm
QT Analysis
QTc formula
Bazett, Fridericia, Framingham, and Hodges
Mindray ECG algorithm Mortara ECG algorithm
QT measurement range
200, 800 ms 300, 600 ms
QT accuracy
200, 800 ms: ± 30 ms
Beyond this range: Not
specified
300, 600 ms: ± 30 ms
Beyond this range: Not
specified
QT resolution
200, 800 ms: 4 ms
Beyond this range: Not
specified
300, 600 ms: 2 ms
Beyond this range: Not
specified
QTc measurement range
200, 800 ms 300, 600 ms
QTc resolution
200, 800 ms: 1 ms
Beyond this range: Not
specified
300, 600 ms: 1 ms
Beyond this range: Not
specified
QT-HR measurement range
Adult: 15, 150 bpm
Pediatric: 15, 180 bpm
N/A