Product Specifications Measurement Specifications
TMS60 Operator’s Manual A - 11
Defibrillation energy absorption
≤ 10% (100  load)
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a
PACE marker:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2 ms
Rise time: 10 s to 100 s
Amplitude: ≥ 0.2mV RTI
Pace pulse rejection
When tested in accordance with the IEC60601-2-27: 201.12.1.101.13,
the heart rate meter rejects all pulses meeting the following
conditions.
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2 ms
Rise time: 10 s to 100 s
About 50% pulse of ANSI/AAMI EC13 (5d) can trigger the pulse
detector, the least changing rate is 20 V/s RTI.
HR
Measurement range
Adult: 0, 15 bpm to 300 bpm
Pediatric: 15 bpm to 350 bpm
Resolution
1 bpm
Accuracy
± 1 bpm or ± 1%, whichever is greater.
Sensitivity
Mindray ECG algorithm Mortara ECG algorithm
200 V 160 V to 480 V