User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- C FCC Compliance
- D Symbols and Abbreviations
- E Anomaly
Product Specifications Mindray Patient Area Network (MPAN) Specification
TMS60 Operator’s Manual A - 9
A.7 Mindray Patient Area Network (MPAN) Specifica-
tion
A.7.1 Technical Specification
A.7.2 Implemented Function
The function implemented by the MPAN is transferring the configuration between the
transmitters.
A.7.3 Performance Specification
Modulation mode
GFSK
Operating frequency
2402MHz to 2480MHz
Channel spacing
2 MHz
Wireless baud rate
1 Mbps
Output power
≤ 2.5 mW
Data Security
128 bit AES
Data integrity
Bit error rate: ≤ 1%
Transmission distance
Distinct vision distance between the transmitters is greater than
or equal to 3 meters.
System capacity
One transmitter can transport configuration information
between at least five other transmitters at the same time in 10
2
room. .
Resistance to wireless interference
The wireless functions of the transmitter are normal when the
following conditions exist simultaneously:
■ The distance between interfering devices (microwave
ovens and Wi-Fi devices) and the transmitter is greater
than 100 centimeters.
■ The distance between interfering devices (cellular
mobile communication network devices and cordless
phones) and the transmitter is greater than is 20
centimeters.
Communication interruption alarm
When communication between transmitters is abnormal, a
corresponding prompt message is displayed on the transmitter