User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- C FCC Compliance
- D Symbols and Abbreviations
- E Anomaly
Product Specifications Power Supply Specifications
TMS60 Operator’s Manual A - 3
A.3 Power Supply Specifications
A.3.1 TD60
The TD60 is powered by batteries.
A.3.2 RC60
The telemetry receiver is powered by external AC power.
Battery type
AA batteries (two or three)
Rechargeable lithium-ion battery (one)
Capacity for lithium-ion
battery
≥ 3500 mAh
Safety and authentication
for lithium-ion battery
UL2054, IEC62133, UN38.3
Run time for (at the
temperature 25 °C ±5 °C, the
screen display is off, no
sound and light as default,
no alarms, at least including
15 minutes operating time
when the screen display is
on, AA battery with PN 0146-
00-0077-10 are used)
Two AA
batteries
Three AA
batteries
Lithium-ion
battery
5-lead ECG: ≥ 48 hours ≥ 72 hours ≥ 72 hours
5-lead ECG +
Nonin SpO
2
≥ 36 hours ≥ 48 hours ≥ 48 hours
5-lead ECG +
Masimo SpO
2
≥ 36 hours ≥ 48 hours ≥ 48 hours
Power On/Off
The startup logo displaying time is less than or equal to three seconds. The
time for entering the main screen to display the waveforms is less than or
equal to 15 seconds.
Shutdown delay
at least 15 minutes after the low battery alarm first occurs
at least five minutes after the critically low battery alarm first occurs
Input voltage
100 to 240 VAC (± 10%)
Frequency
50 Hz/60 Hz (± 3 Hz)
Input current
0.6 A to 0.3 A