User's Manual

Table Of Contents
Maintenance Verifying the ECG at the TD60
TMS60 Operator’s Manual 14 - 5
A square wave appears on the screen and the message ECG Calibrating is dis-
played.
6. Compare the amplitude of the square wave with the wave scale. The difference
should be within 5%.
7. After completing the verification, disable Calibrate ECG.