User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- C FCC Compliance
- D Symbols and Abbreviations
- E Anomaly
Physiological Alarm Messages at the CS Troubleshooting
12 - 4 TMS60 Operator’s Manual
ECG ECG Weak Signal High The ECG signal is so weak
that the monitor can’t
perform ECG analysis.
Check the patient’s
condition and the ECG
connections.
Asystole High The patient is in
arrhythmia.
Check the patient’s
condition and the ECG
connections.
Vfib/Vtac High
VTac High
Vent. Brady High
Extreme Tachy High
Extreme Brady High
R on T Medium*
VT > 2 Low*
Couplet Message*
PVC Message*
PVCs/min Medium*
Bigeminy Medium*
Trigeminy Medium*
Tachy Medium*
Brady Medium*
Missed Beat Message*
Vent. Rhythm Medium*
Pause Medium*
PNP (Pacer Not
Pacing)
Message* The pacer appears
abnormal.
Check the pacer.
PNC (Pacer Not
Capture)
Message*
Measurement Alarm message Alarm level Possible cause Solution
* The asterisk (*) means the alarm level is configurable.