Maintenance Menu Configuring the TD60
10 - 4 TMS60 Operator’s Manual
3. Tap to exit the Alarms menu.
Style Allows an authorized user to set the alarm tone
pattern.
ISO, Mode 1, Mode 2
Timeout
Reminder Tone Allows an authorized user to enable or disable
the reminder tone.
If the alarm tone is turned off, enabling this set-
ting can issue a periodic reminder tone.
On, Off
Reminder Interval Allows an authorized user to configure the
intervals between the alarm tones.
1 min, 2 min, 3 min, 5 min, 10 min
Minimum Alarm Volume
Technical Allows an authorized user to set the minimum
technical alarm volume.
The minimum technical alarm volume defines
the minimum value you can set for the techni-
cal alarm volume.
For example:
If the minimum technical alarm volume is set to
5, the minimum value you can set for the tech-
nical alarm volume in the Audio Volume menu
is 5 (as shown in the following figure).
If the minimum technical alarm volume is set to
Off, and the technical alarm volume can be set
to 0, the alarm sound is turned off and the
symbol appears on the screen.
Off, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10
Technical Alarm Priority
ECG Lead Off Allows an authorized user to configure the
alarm level.
Low, Medium, High
SpO
2
Sensor Off Allows an authorized user to configure the
alarm level.
Low, Medium, High
Section & Options Description Settings*
* The factory default settings are in bold.
Minimum value