9 - 30 TMS60 Operator’s Manual
9.6.3 Arrhythmia Settings
The following table lists the arrhythmia settings in the Arrhythmia Analysis section of
the ECG tab.
9.6.4 Relearning
A relearn can be done for arrhythmia, ST analysis, or for both simultaneously.
The TD60 initiates the learning process for ST or Arrhythmia analysis after any of the fol-
lowing:
TD60 power-up
Return to normal monitoring from the Standby mode
Enabling ST or Arrhythmia analysis
The lead has been changed in ECG 1 waveform (3 lead only)
Patient size changes
Selecting the Relearn button in the Arrhythmia Analysis section
Selecting the Relearn button in the Arrhythmia Analysis section is recommended after
one or more of the following:
ECG electrodes have been repositioned
Eight hours have passed since the last relearn
Significant changes occurred to the patient QRS complex
Options or Buttons Description Settings*
Arrhythmia Alarms Configures the arrhythmia alarm settings. Refer to "Factory Default Arrhyth-
mia Alarm Settings" on page 9 - 6
for details.
Relearn
Enables an
arrhythmia relearning.
Refer to "Relearning" on page 9 -
30 for details.
QRS Threshold Settings Configures the QRS threshold. Refer to "Configuring the QRS
Threshold" on page 9 - 31 for
details.
NOTE
ST Analysis must be turned on before it will relearn.