Monitoring with the TD60 at the CS ST Monitoring
TMS60 Operator’s Manual 9 - 25
3. Select the OK or Cancel button.
The OK button saves the settings and closes the menu.
The Cancel button closes the menu without saving the settings.
9.5.4 ST Display
When ST Analysis is enabled, the ST digital area displays:
1. Area name
2. Activation State Off icon for ST alarm
3. ST values
4. Units of measure
5. Lead identifier
Example 5-lead ST digital area
ISO/J Selecting Auto fixes the ISO and J/ST
points.
Selecting Manual allows the clinician to
manually adjust ISO and J/ST points.
Auto, Manual
ISO left arrow If ISO/J is set to manual, the button adjusts the ISO reference line to the left.
ISO right arrow If ISO/J is set to manual, the button adjusts the ISO reference line to the right.
J left arrow If ISO/J is set to manual, the button adjusts the J reference line to the left.
J right arrow If ISO/J is set to manual, the button adjusts the J reference line to the right.
Buttons Description Settings*
* The factory default settings are in bold.
1
2
4
5
3