Monitoring with the TD60 at the CS ECG Monitoring
TMS60 Operator’s Manual 9 - 15
3. Click Exit to save the settings and close the Waveform Setup menu.
9.3.3 Other Settings
The following table lists all settings in the Other Settings menu.
ECG 2 Select the desired ECG lead and set the
corresponding gain for ECG 2 to display
in the waveform area.
I, II, III, aVR, aVL,
aVF, V
× 0.125, × 0.25, × 0.5,
× 1, × 2, × 4
ECG 3 Select the desired ECG lead and set the
corresponding gain for ECG 3 to display
in the waveform area.
I, II, III, aVR, aVL,
aVF, V
× 0.125, × 0.25, × 0.5,
× 1, × 2, × 4
Primary lead Configures the primary analysis lead. I, II, III, aVR, aVL,
aVF, V
None
Secondary lead Configures the secondary analysis lead. I, II, III, aVR, aVL,
aVF, V
None
Classification lead
(only for Mortara ECG
algorithm)
Configures other analysis lead. I, II, III, aVR, aVL,
aVF, V
None
Options Description
Settings*
Lead Waveform size
* The factory default settings are in bold.
Options Description Settings*
Paced The option is unavailable at the CS. The paced status can be set at the TD60 only.
Refer to "Checking the Paced Status" on page 7 - 10 for details.
Pacer Reject Configures whether or not to reject the pace
pulses.
On: the pace pulses are not counted as
extra QRS complexes.
Off: the pace pulses are not rejected.
On, Off
This Pacer Reject option is only available when Paced is set to Ye s at the TD60.
Filter Configures the ECG filter in all operating
modes.
Monitor: use under normal
measurement conditions.
ST: use when the ST monitoring is
applied.
Monitor, ST
* The factory default settings are in bold.