ECG Monitoring Monitoring with the TD60 at the CS
9 - 12 TMS60 Operator’s Manual
9.3 ECG Monitoring
At the CS, you can view and change the heart rate (HR), QT, ST, and arrhythmia settings in
the ECG tab of the Parameter Setup menu.
For details about the Parameter Setup menu, refer to the BeneVision Central Station
Operator’s Manual (P/N 046-007960-00).
9.3.1 HR Settings
The following table lists the HR settings in the HR section of the ECG tab.
Options or Buttons Description Settings*
HR Activation State Configures whether or not to enable the
HR alarm.
On, Off
HR Alarm Priority Configures the HR alarm levels. High, Med
HR High Limit (bpm) Configures the HR high alarm limit. Adult:
(Low limit + 2) to 300
The default is 120.
Pediatric:
(Low limit + 2) to 350
The default is 160.
HR Low Limit (bpm) Configures the HR low alarm limit. Adult:
15 to (high limit - 2)
The default is 50.
Pediatric:
15 to (high limit - 2)
The default is 75.
Pacer Rate (bpm) (only
Mortara ECG algorithm)
When some pacemaker pulses are difficult
to reject, the pulses are counted as a QRS
complex and could result in an incorrect
HR and failure to detect some arrhythmias.
Configures the pacemaker rate, the sys-
tem can calculate HR and detect arrhyth-
mias more accurately.
40 to 100
The default is 60.
The option is unavailable when Paced is set to No at the TD60. For details about
the paced status, refer to "Checking the Paced Status" on page 7 - 10.
Sweep Speed Configures the ECG wave speed. 6.25 mm/s, 12.5 mm/s, 25 mm/s,
50 mm/s
HR Source Configures the HR source. ECG, SpO
2
, Auto, Both
* The factory default settings are in bold.