User's Manual

B-3
Surge IEC
61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not apply
Voltag
e dips, short
interruptions a
nd
voltage var
iations
on pow
er supply
input lines IE
C
61000-4-11
<5 % U
T
(>95 % dip in
U
T
) for 0.5 c
ycle
40 % U
T
(60 %
dip in U
T
)
for 5 c
ycles
70 % U
T
(30 %
dip in U
T
)
for 25 c
ycles
<5 % U
T
(>95 % dip in
U
T
) for 5 s
Not apply
/
Pow
er frequency
(50/60 HZ)
magnetic fie
ld IEC
61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
N
ote: U
T
is the AC ma
ins voltage prior to application of the test level.
Guidance and Declarati
on - Electromagnetic Immunity
The dev
ice is intended for use in the specified electromagnetic environment. The customer or the user of the
device should assure that it is used in such an environment as described below.
Immun
ity test
IEC60601
test level
Compliance
level
Electromagnetic env
ironment -
guidance
Cond
uced RF
IEC610
00-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any part
of the system, including cables, than the
recommended separation distance calculated
from the equation appropriate for the frequency
of the transmitter. Recommended separation
distances:
Pd 2.1
Rad
iated RF
IEC610
00-4-3
3V/m
80MHz to
2.5GHz
3V/m Recommended separation distances:
80 MHz800 MHz
Pd 2.1
800MHz-2.5GHz