User's Manual
B-2
RF em
issions CISPR 11 Class B The device is suitable for use in all establishments ,
domestic and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes
Harmon
ic emissions
IEC610
00-3-2
Not apply /
Voltag
e Fluctuations/Flicker
Emiss
ions IEC 61000-3-3
Not apply
If the s
ystem is operated within the electromagnetic environment listed in Table Guidance and
Declaration —Electromagnetic Immunity, the system will remain safe and provide the following
essential performance,
Operating mode
Accuracy
Function
Accessories identification
Data stored
Alarm
Detect for connection
Guidance and Declarati
on - Electromagnetic Immunity
The dev
ice is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment.
Immun
ity test
IEC60601 te
st level
Compliance level
Electromagnetic
environment - guidance
E
lectrostatic
disch
arge (ESD)
IEC 6100
0-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
E
lectrical fast
transient/burst IEC
61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
Mainspo
werqualityshould
bethatofatypical
commercialorhospital