User's Manual

21
Figure 13. Parameter setting interface of manual calibration
Note:
Temperature measurement can be started till 5~10 minutes after the boot when the
measurement accuracy has already been stabilized.
The calibration parameter has a great influence on the measurement precision,
so the calibration parameter should be reset when the precision has a large error
during the usage;
During the whole calibration process, the center of the screen must be targeted at
the uncovered forehead or ears of the assisting person. Otherwise, the calibration
result may leave a large margin error.
During the calibration process, the calibration assistant person needs to stand still
until the calibration is completed.
If a large margin error of the temperature measurement occurs as a result of
maloperation during the calibration process, please re-execute the one-click
calibration or manual calibration procedure.
4.2.5. Selection of Color Palette
Two color palettes Rainbowand Iron Grayare available for the
users to choose on the Parameter setting page according to their
preference.