Temperature measurement can be started till 5~10 minutes after the boot when the
measurement accuracy has already been stabilized.
When the thermal imaging module is calibrated, it clicks every few seconds, which
may be accompanied by a short-time video lag.
During the whole calibration process, the center of the screen must be targeted at
the uncovered forehead or ears of the assisting person. Otherwise, the calibration
result may leave a large margin error.
During the calibration process, the calibration assistant person needs to stand still
until the calibration is completed.
If a large margin error of the temperature measurement occurs as a result of
maloperation during the calibration process, please re-execute the one-click
calibration or manual calibration procedure (see below).
4.2.4. Manual Calibration
The helmet is ready for temperature measurement after One-Click
calibration. However, manual calibration can be applied to adjust the
temperature parameters to get a higher accuracy. Manual calibration
requires the assistance of a second person of normal body temperature.
Step 1: Let the assisting person stand 1 meter from the thermal
imager and measure his forehead temperature with the thermal
imager;
Step 2: The temperature reading will rise as increasing the
temperature adjusting parameter (see the red frame below) and vice
versa. The temperature adjusting parameter should be within the
range of -10~+10.
Step 3: Click "save" to confirm the parameters.