Guidance and manufacture’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of device
should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Not applicable Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0 % U
T
; 0.5 cycle
at 0
°
,45
°
,90
°
, 135
°
,
180
°
, 225
°
, 270
°
, 315
°
0 % U
T ;
1 cycle
70 % U
T
; 25/30 cycle
0% U
T
; 250/300 cycle
Not applicable Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the
device requires continued
operation during power mains
interruptions, it is recommended
that the
device be powered from
an uninterruptible power supply or
a battery.
Power frequency
(50Hz/60Hz)
magnetic field IEC
61000-4-8
30 A/m
50/60Hz
30 A/m
50/60Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.