Operator's Manual
6-2
To select the lead mode, follow the steps below:
1. In the main screen, click [ ] → [Setup] to enter the setting
screen.
2. Click [ECG Setup] → [Lead Mode] to set the required lead mode.
3. Return to the main screen after setting.
6.3 Setting ECG Waveform and Report
Set the ECG waveform and report before starting an ECG
measurement. Operation procedures:
1. Click the quick keys at the bottom of the main screen to set the
speed, sensitivity, display format and filter frequency
respectively.
2. Click [Setup] → [ECG Setup] and [Record Setup] to check other
waveform setting items and report setting items, and make
relevant settings as needed.
For more information, see
Chapter 8 System Settings
.
6.4 Acquisition and Analysis
6.4.1 Auto Diagnosis
In the local login mode, after the ECG waveform is stable, click the
[Start] button, the device starts recording the ECG waveform. After
the ECG data is acquired for the set time period, the device
automatically starts analysis, and selects whether to print ECG
report according to the settings.