Warning
During defibrillation, use disposable electrodes and ECG adapter wires
specified by the manufacturer and use them in accordance with their
instructions for use.
Warning
After defibrillation, the ADS filter is set at 0.67Hz, and the cardiogram is
displayed and maintained within 10 seconds.
Warning
Use only the patient cable and electrodes supplied by the manufacturer
while defibrillating.
Warning
For patient with a pacemaker, since this device has a pacing signal
suppression function, under normal circumstances, pacing pulses will not
be included in the pulse rate detection and calculation. However, if the
width of the pacing pulse exceeds 2ms, it is still possible to continue
counting the pacing pulse. To reduce this possibility, the operator should
closely observe the changes in the ECG waveform on the screen, and do
not rely on the indications of the device itself, when the device is used for
such patients.
1.1.3 Battery Warnings
Warning
Improper operation may cause the lithium battery (hereinafter called
battery) to be hot, ignited or exploded, and it may lead to the decrease of
the battery capacity. It is necessary to read this manual carefully and pay
more attention to warning information.