Operator's Manual
B-5
Table 3
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Immunity
Test
IEC 60601 Test Level
Compliance Level
Conducted
RF
IEC
61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM bands between 0.15
MHz and 80 MHz
3 V
0.15 MHz to 80 MHz
6 V in ISM bands between 0.15
MHz and 80 MHz
Radiated RF
IEC
61000-4-3
10V/m
80 MHz to 2.7 GHz
10V/m
Table 4
Guidance and Manufacturer’s Declaration - IMMUNITY to proximity fields from RF
wireless communications equipment
Immunity
Test
IEC60601 Test Level
Compliance
Level
Test
Frequency
Modulation
Maximum
Power
Immunity
Level
Radiated
RF
IEC
61000-4-3
385 MHz
**Pulse
Modulation:
18Hz
1.8W
27 V/m
27 V/m
450 MHz
*FM+ 5Hz
deviation:
1kHz sine
2 W
28 V/m
28 V/m
710 MHz
745 MHz
780 MHz
**Pulse
Modulation:
217Hz
0.2 W
9 V/m
9 V/m