Operator's Manual
B-4
Table 2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic
discharge (ESD) IEC
61000-4-2
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV
air
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15
kV air
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Surge IEC61000-4-5
± 0.5kV, ± 1 kV line(s) to lines
± 0.5kV, ± 1 kV, ± 2 kV line(s)
to earth
± 0.5kV, ± 1 kV line(s) to
lines
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
0 % UT; 0.5 cycle At 0°, 45°,
90°, 135°, 180°, 225°,
270°and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
0 % UT; 0.5 cycle At 0°,
45°, 90°, 135°, 180°,
225°, 270°and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.