Operator's Manual
B-2
Warning
The device may still be interfered even if other devices meet the emission
requirements of the corresponding national standards.
Caution
Users shall install and use the device according to the EMC information
provided in the this manual.
Caution
Mobile or portable RF communication equipment may affect the
performance of the device. Avoid strong electromagnetic interference
when in use, such as near mobile phones, microwave ovens, etc.
Caution
When the input signal amplitude is lower than the minimum amplitude
(20μVp-p) specified in the technical specifications, the measurement result
may be inaccurate.
Caution
The customer or the user of the device should assure that the device is
used under the electromagnetic environment specified below, otherwise
the device may not work normally.
The following cables must be used to meet electromagnetic
emission and anti-interference requirements:
No.
Name
Length
Shield (Yes/No)
1
Patient cable
Approximately 3.2m
Yes
2
DC power adapter
cable
1.2m
No
3
Power cable
1.8m
No