Operator's Manual

10-2
10.2.2 Cleaning Patient Cable and Electrodes
Before cleaning the patient cable and electrodes, remove the patient
cable from the device.
For the cleaning of the patient cable and electrodes, refer to their
instructions for use delivered with the accessories.
10.2.3 Cleaning the Thermal Print Head
Dirty print head deteriorates the printing quality. Clean the print head
at least once a month or as needed.
To clean the thermal print head, follow the steps below:
1. Turn off the recorder base.
2. Open the paper compartment cover and take out the recording
paper.
3. Wipe the print head gently with a clean soft cloth dampened in
75% alcohol. For stubborn stain, soak it with a small amount of
alcohol first and wipe it off with a clean soft cloth.
4. Reload the recording paper and close the paper compartment
cover after the print head is completely air dried.
10.3 Disinfection
Disinfection of main unit of the device is not necessary. To avoid
permanent damage to the device, disinfection can be performed only
when it has been considered as necessary according to your
hospitals regulations. Before disinfection, clean the device first.
For the disinfection of the patient cable and electrodes, refer to their
instructions for use delivered with the accessories.