Operator's Manual

10-1
Chapter 10 Cleaning, Disinfection and
Maintenance
Sterilization is not recommended for this device and its accessories,
but they should be kept clean. If the device has become
contaminated, clean it before disinfection.
10.1 Recommended Cleaning Agents
Supported cleaning agents: water, neutral soap solution, ethanol
solution (volume ratio: 70% to 80%).
Supported cleaning tools: cotton ball, soft gauze, soft brush, soft
cloth.
10.2 Cleaning
10.2.1 Cleaning the Device
Clean the exterior surface of the device monthly or more frequently if
needed. Before cleaning the device, consult your hospitals
regulations for cleaning the device.
To clean the device, follow the steps below:
1. Turn off the device and disconnect it from the power cable and
accessories.
2. Clean the surface of the device with a clean soft cloth moistened
with one of the recommended cleaning agents.
3. Wipe off all the cleaning agent residue with a clean dry cloth.
Dry your device in a ventilated, cool place.