Operator's Manual

9-3
No.
Messages or Troubles
Solutions
Nobody is touching the
patient.
There is no powerful electrical
equipment operating nearby,
such as X-ray machines or
ultrasonic instruments.
The patient is not wearing
glass or diamond ornaments.
AC filter frequency is properly
set.
If the interference cannot be
cleared after the above measures,
use an AC filter, and the recorded
waveform is slightly attenuated.
10
EMG Interference
Symptom: The ECG has
irregular fluctuation while the
baseline demonstrates no
change.
Check the following aspects of the
device for solving problems:
The room is uncomfortable?
The patient is nervous or feels
cold?
The bed is too narrow?
The patient is talking?
The limb electrode clamps are
attached too tightly?
If the interference cannot be
cleared after the above measures,
use an EMG filter, and the