Use Instruction
16
No power.
No display appears on the
unit.
Re-charge battery.
Measurement values appear
too high or too low.
Blood pressure varies constantly.
Many factors including stress, time of day, and how you wrap the CUFF,
may affect your blood pressure, where you position your Arm, may affect
your blood pressure.
Review the sections “BLOOD PRESSURE MEASUREMENT METHOD ”.
11.2 Error indicators (Error Code)
SYMBOL
CAUSE
CORRECTION
Err1
Sensor abnormality
When measuring, avoid the environment (e.g. to
magnetic fields, electromagnetic fields, external
electrical influences etc.).
Err2
No pulse or blood pressure reading due
to very weak artery pulse.
Remove CUFF. Wait 2-3 minutes and take another
measurement. Repeat steps listed under the clause
“BLOOD PRESSURE MEASUREMENT METHOD”
Inflate the CUFF 40mmHg higher than the previous
inflation value.
The CUFF is too loose
Apply the CUFF correctly, Refer. to the clause
"Applying the CUFF'’
Clothing is interfering with the CUFF
Remove any closing interfering with the CUFF.
Refer to the clause “Applying the CUFF”
Err4
Air plug disconnected
Insert the plug securely, refer to the clause
"Applying the CUFF'’
Air cuff not applied correctly
Apply the CUFF correctly. Refer to the clause
“Applying the CUFF”
Air is leaking from the CUFF
Replace the CUFF with the new one.
Others
Contact Customer Service.
Err5
Pipe bend causing air blockage
Apply the CUFF correctly. Refer to the clause
“Applying the CUFF”
connector blockage
Replace connector
Others
Contact Customer Service
Err6
The interference is too large when in
measurement
Repeat measurement. Remain still and do not talk
during measurement. Refer to the clause “BLOOD
PRESSURE MEASUREMENT METHOD”.