User's Manual

Thermometer eTE-89B4 User Manual V1.0
16
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
N/A
be that of a
typical commercial or
hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % UT; 0,5 cycle g) At
, 45°, 90°, 135°, 180°,
225°, 27 and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
N/A
Mains power quality should
be that of a
typical commercial or
hospital environment. If the
user of the Model + Name
requires continued operation
during power mains
interruptions, it is
recommended that the
Model + Name be powered
from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE U
T
is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer´s declaration electromagnetic immunity
for EQUIPMENT and SYSTEM
Guidance and manufacturer´s declaration electromagnetic immunity
The Model + Name is intended for use in the electromagnetic environment specified below. The customer
or the user of the Model + Name should assure that it is used in such an environment.
Immunity
test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Model + Name, including cables,
than the recommended separation distance
calculated from the equation applicable to the