User's Manual
Thermometer eTE-89B4 User Manual V1.0
15
Voltage fluctuations
flicker emissions
IEC 61000-3-3
N/A
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration –
electromagnetic immunity
The Model + Name is intended for use in the electromagnetic environment specified below. The customer
or the user of the Model + Name should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
Floors should be wood,
concrete or
ceramic tile. If floors are
covered with
synthetic material, the
relative humidity
should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
N/A
Mains power quality should
be that of a
typical commercial or
hospital
environment.
Surge
Mains power quality should