User's Manual

17 18
Instructions for use
Technical description
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments
and so on.
WarningDon’t near active HF surgical equipment and the RF shielded room of an
ME system for magnetic resonance imaging, where the intensity of EM disturbances
is high.
WarningUse of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that
they are operating normally.
WarningUse of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.”
WarningPortable RF communications equipment (including peripherals such as
antenna cables and
external antennas) should be used no closer than 30 cm (12
inches) to any part of the equipment (detail model name), including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
If anya list of all cables and maximum lengths of cables (if applicable),
transducers and other ACCESSORIES that are replaceable by the RESPONSIBLE
ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT
or ME SYSTEM with the requirements of Clause 7 (EMISSIONS) and Clause
8 (IMMUNITY). ACCESSORIES may be specified either generically (e.g.
shielded cable, load impedance) or specifically (e.g. by MANUFACTURER and
EQUIPMENT OR TYPE REFERENCE).
1.all necessary instructions for maintaining BASIC SAFETY and ESSENTIAL
PERFORMANCE with regard to electromagnetic disturbances for the excepted
service life.
2. Guidance and manufacturer’s declaration -electromagnetic emissions and
Immunity
If anythe performance of the ME EQUIPMENT or ME SYSTEM that was
determined to be ESSENTIAL PERFORMANCE and a description of what the
OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or
degraded due to EM DISTURBANCES (the defined term ESSENTIAL
PERFORMANCE” need not be used).
Table 1
Guidance and manufacturers declaration - electromagnetic emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Emissions test Compliance
Group 1
Class B
Class A
Compliance
and, if not installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
Consult the dealer or an experienced radio/TV technician for help.
The rating information is located at the bottom of the unit.